Pratt & Whitney LMU Bearing Measuring Instrument
|
|
|
The
Labmaster Universal bearing measuring instrument can
measure a wide range of anti-friction bearings including
tapered bearings, needle bearings, ball bearings,
spherical bearings, and cylindrical bearings.
A few of the most popular bearing characteristics that
can be measured on the Pratt & Whitney Labmaster bearing
measuring system include:
|
-
Radial
Internal Clearance (Radial Internal Play)
-
Axial Internal Clearance
(Axial Internal Play / End Play)
-
Internal Diameter of Bearing (Inner Ring)
-
External Diameter of Bearing (Outer Ring)
-
Inner Raceway of Bearing
-
Outer Raceway of Bearing
-
Shield Thickness
-
Ball Bearing Diameter
|
 |
|
|
|
|
Easy
to use and the most accurate bearing measuring
instrument on the market today, the P&W Labmaster
Universal measures bearings used for the aerospace,
automotive, and medical industry. A popular add-on for
the bearing remanufacture & repair industry is the axial
internal clearance (axial play) and radial internal
clearance (radial play) kits. These accessories can
measure bearing wear and provide important failure
analysis.
|
|
- Differential Measurement
Eliminates errors due to deflections of
machine elements and errors due to variations in
mechanical connections.
- Floating Ring Design
Flexure support allows 6 degrees of
frictionless movement with zero looseness. The
force application system is also frictionless.
The absence of friction and looseness ensures
repeatability. The floating ring design allows
precise alignment of the bearing races, which
reduces misalignment errors.
- Horizontal Orientation
Minimizes errors due to the bearing
deforming under its own weight. Eliminates the
need for the force system to compensate for
bearing weight.
- Concentrated Force Application
Minimizes error due to ring deflection.
Radial Play Kit includes force systems (load
presets) to satisfy ABMA, ANSI, ISO, JIS, and
DIN standards for ABEC ball bearings, RBEC
roller bearings, etc.
- Performance Highlights
Highest accuracy, repeatability, and
resolution on the market.
|
|
|
"Other" Gage
|
P&W Radial
Internal Clearance Kit |
| Inner ring mounted on 3 Jaw chuck.
If chuck is too tight then ring will deform. If
chuck is too loose then gap will affect measurement. |
No radial loads are applied when
mounting the bearing. |
| Single ended measurement will
include errors due to machine element deflections,
fixturing looseness, and air bearing variation. |
Differential measurement is
unaffected by machine element deflections, fixturing
looseness, or air bearing variation. |
Large offset distance between the
force application and reaction points causes bending
of the bearing rings, which will add or subtract
from the measurement.
|
Minimal offset distance (< 0.125")
between the force application and reaction points
ensures minimal flexing of the bearing. A
0.125"offset will have 500X less deflection than a
1" offset. |
| No provision for precisely aligning
the bearing races. Alignment will be affected by
machine level and force application friction. |
The floating ring design allows
precise alignment of the bearing races, minimizing
errors due to misalignment. |
| Friction in the force system due to
cylinder seals causes force variation. |
Frictionless force application
minimizes force variation. |
| Friction between the force
application V and the outer ring affects outer ring
centering. |
Frictionless force application
allows the outer ring to auto-center. |
| Vertical mounting of the bearing
allows the ring to deform under its own weight,
directly affecting the measurement. |
Horizontal mounting of the bearing
minimizes ring deflection due to its own weight.
Ring deflection is in a direction that does not
directly affect the measurement. |
| Vertical mounting of the bearing
means that the weight of the outer ring will add to
the downward applied force and subtract from the
upward applied force. |
Horizontal mounting of the bearing
ensures that the bearing load is unaffected by the
outer ring weight. |
|
|
More Dimensional
Measuring MachinesTop |
|
|
| |
Pratt
& Whitney Measurement Systems, Inc.
66 Douglas Street, Bloomfield, CT 06002-3619
TEL: (860) 286-8181 or 1-800-371-7174 FAX: (860)
286-7878
© Pratt & Whitney Measurement Systems, Inc. |
|